| Model Number |
380601-45 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse went on site, reproduced the issue and replaced the single-port one axis manipulator controller board (soam) to resolve the error.The system was tested and verified as ready for use.Isi received a da vinci product involved with this complaint to perform failure analysis.\
\.
|
| |
|
Event or Problem Description
|
|
It was reported that user informed the technical support engineer (tse) that error 32100 remained active and did not recover even after an emergency power off was performed.The tse reviewed the error log and determined it pointed to single-port one axis manipulator (soam) related to the entry guide manipulator (egm) pitch control hardware on the left-side (soam) board.There was no report of patient involvement or patient injury.
|
| |
|