Intuitive surgical, inc.(isi) did receive the harmonic ace instrument to perform failure analysis (fa).Fa was able to confirm/reproduce the reported complaint.The instrument was found to have curved blade broken at its base.A piece approximately 11.0mm x 2.1mm was found to be broken off and was returned with the instrument.Components adjacent to this broken blade do not show damage.The probable root cause is attributed to use conditions.Broken blades result from blade scratches and damage caused by contact with other instruments or hard metal objects (e.G., staples, clips, etc.) while the instrument is activated.These initial damages can worsen due to the ultrasonic vibrations of the harmonic ace blade, leading to blade failure.Blade damage may be detected by the generator with a solid tone or an error.
|