Product analysis the device was returned with no red safety tab mechanism and in a deployed state.No stent was returned with the device.The guidewire used during the procedure stuck along with the device, the device was identified as 80mm by the strain relief the device handle was opened, and the pull cable was attached to the thumbwheel scroll but was detached from the device, the thickness of the stuck guidewire using the laser micrometre was measured as 0.014¿ instead of 0.035¿ which indicates that the incorrect guidewire size was used, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|