| Model Number |
PVS23 |
| Medical Device Problem Code |
Collapse (1099)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The manufacturer was notified of the intraoperative explant of a perceval s size m valve.The following provides a comprehensive summary of all information currently available to the manufacturer.On (b)(6) 2026, an attempt was made to implant a perceval s valve, size m (pvs23).However, stent infolding was observed during the procedure.Consequently, the valve was explanted and replaced with another perceval valve of unknown size.No further information is available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturing and material records for the device involved in this event, as they pertain to the reported issue, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this prosthesis satisfied all required material, visual, and performance standards at the time of manufacture and release.The manufacturer will submit follow-up report upon availability of new details from healthcare facility and/or completion of device investigation.
|
| |
|