The account reported finding chg leaked in pack.A photo was received for the investigation, no physical sample was available, and the brown paper tray appeared wet and moist, which clearly confirms that fluid has leaked inside the pack.Based on the details and photo received, a leak was found inside the pack; however, since no physical sample was available to examine, the true root cause could not be accurately determined at this time.The division will monitor this issue for trending purposes, and a risk assessment has been completed and the incident was found to be acceptable.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|