The account reported finding open seal.Finished good production records were reviewed, and there were no line downs, deviations or non-conformance reports that might have had an impact on this issue; finished good pack build was reviewed, and there are no issues with the build that likely contributed to the reported issue.A physical sample for one pack was received, and the tyvek seal was observed open along the top middle of the bag, confirming the reported issue; a white line running along the edge seam indicated that the pouch did go through the sealing machine and was closed at some point in time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|