| Catalog Number |
SRM-310 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "exchange with no complaint against the device".Later healthcare professional reported "fractured implant" which was discovered during surgery.The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "exchange with no complaint against the device".Later healthcare professional reported "fractured implant".Later healthcare professional "neither implant was ruptured" and "broken gel".This record is for the left side.Device has been explanted.Singulair was prescribed for contralateral side capsular contracture.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Continued e1 -- alternate phone number:(b)(6).Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: removal due to contralateral side capsular contracture and rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d9, h3, h6.Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- rupture: observed an opening through microscopic inspection assessed as surgical damage.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|