BOSTON SCIENTIFIC NEUROMODULATION CORPORATION Infinion? CX; stimulator, spinal-cord, totally implanted for pain relief
|
Back to Search Results |
|
| Model Number |
SC-2317-50 |
| Medical Device Problem Codes |
High impedance (1291); Unexpected Therapeutic Results (1631)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
03/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced inadequate stimulation and charging issues due to high impedances on the right lead.Program optimization was attempted and was unsuccessful.As a result, the patient underwent a revision procedure wherein the spinal cord stimulator (scs) lead and implantable pulse generator (ipg) were replaced.The patient was reported to be doing well postoperatively.The explanted devices were not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced inadequate stimulation and charging issues due to high impedances on the right lead.Program optimization was attempted and was unsuccessful.As a result, the patient underwent a revision procedure wherein the spinal cord stimulator (scs) lead and implantable pulse generator (ipg) were replaced.The patient was reported to be doing well postoperatively.The explanted devices were not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: exact date unknown, event occurred in (b)(6) 2026.Block d.2b; additional applicable product code: qrb.Additional suspect medical device components involved in the event:
model number/catalog number: sc-1160.Serial number: (b)(6).Batch/lot number: (b)(6).Model/catalog description: spectra wavewriter ipg kit.Unique identifier (udi) #: (b)(4).Model number/catalog number: sc-4318.Serial number: n/a.Batch/lot number: 22420722.Model/catalog description: clik x anchor sterile kit.Unique identifier (udi) #: (b)(4).Investigation results
the implantable pulse generator (ipg), spinal cord stimulator (scs) lead, and anchor were not returned for analysis; therefore, a technical analysis could not be performed.Device history record
it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Media review
review of the provided image confirmed the reported issue of high impedance.No additional observations were noted in the media.Similar complaint review
a review of complaints with similar reported conditions was completed.The assessment did not identify any emerging trends, meaningful commonalities, or similar complaints requiring further escalation.Based on the review results, no adverse trend was identified; therefore, no additional escalation or action is required at this time.Investigation conclusion:
based on a thorough review of the reported complaint, an investigation conclusion code of cause not established was assigned to the investigation for all components.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: exact date unknown, event occurred in march 2026.Block d.2b; additional applicable product code: qrb.Additional suspect medical device components involved in the event:
model number/catalog number: sc-1160,
serial number: (b)(6),
batch/lot number: 340488,
model/catalog description: spectra wavewriter ipg kit,
unique identifier (udi) #: (b)(4).Model number/catalog number: sc-4318,
batch/lot number: 22420722,
model/catalog description: clik x anchor sterile kit,
unique identifier (udi) #: (b)(4).Investigation results,
the implantable pulse generator (ipg), spinal cord stimulator (scs) lead, and anchor were not returned for analysis; therefore, a technical analysis could not be performed.Device history record,
it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review,
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Media review
review of the provided image confirmed the reported issue of high impedance.No additional observations were noted in the media.Similar complaint review,
a review of complaints with similar reported conditions was completed.The assessment did not identify any emerging trends, meaningful commonalities, or similar complaints requiring further escalation.Based on the review results, no adverse trend was identified; therefore, no additional escalation or action is required at this time.Investigation conclusion:
based on a thorough review of the reported complaint, an investigation conclusion code of cause not established was assigned to the investigation for all components.
|
| |
|