• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; BED ONLY,4-MOTOR, HI/LOW,34.5W, LOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; BED ONLY,4-MOTOR, HI/LOW,34.5W, LOCK Back to Search Results
Catalog Number FCE1385B
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the section of the bed where the head is welded into the frame is broken.Additionally, the customer states that the control boxes for both beds are only raising the foot portion of the bed.".
 
Additional Manufacturer Narrative
It was reported that they have two beds with the same issue.The section of the bed where the head is welded into the frame is broken.The control boxes for both beds are only raising the foot portion of the bed.It was reported that they have two beds with the same issue.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available, this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
BED ONLY,4-MOTOR, HI/LOW,34.5W, LOCK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
dantchi jeliazkova
three lakes drive
northfield, IL 60093--275
MDR Report Key50235205
Report Number1417592-2026-16650
Device Sequence Number13330400
Product Code FNL
UDI-Device Identifier10888277100022
UDI-Public10888277100022
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 08/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberFCE1385B
Device Lot NumberFCET2017020027
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/01/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-