(b)(4)
date sent 8/31/2026.D4 batch # 947d93.Corrected data d4: udi#.Investigation summary.The product was returned for thorough evaluation, which included visual inspections of the returned device.Visual analysis of the returned sample determined that the cdh25p device was received with no apparent damage and the anvil was not returned for analysis.Only casing half washer was returned and there were no staples present.When the test battery was installed was noted difficulty to insert the battery and the green checkmark illuminated, indicating that the device was fired and achieved complete firing stroke.The device was disassembled to verify the condition of the internal components and the bulkhead contact from the left side was noted to be damaged.In addition, marks and damages on the bulkhead shroud were noted possibly caused by improper battery loading.This condition may have caused the reported event.Although no conclusion could be reach on the cause of the reported event, it should be noted that when insert the battery pack, aligning the release tabs with the slots situated on the back of the device.Ensure that it is completely engaged; a clear click sound will confirm it¿s properly secured, and the backlight of the tissue compression scale will glow.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 947d93, and no non-conformances related to the reported complaint condition were identified.
|