| Model Number |
PFA-GEN-01 |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise on thermal damage found during analysis.During an af procedure using rf and pfa (tactiflex duo), it was noted that the following error message was displayed on the current generator; "hard drive issue".The tactiflex duo catheter was replaced but the issue persisted.The procedure was completed using the tacticath.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
One current pfa generator (sn: (b)(6)) was received for evaluation.The reported event was confirmed through a post failure that identified an internal load test failure.Based on the information provided to abbott and the investigation performed, the root cause of the post failure can be attributed to thermal damage to the pinnacle bfc pca.The batch record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|
| |
|