• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Katalyst Surgical, LLC 27ga DEX¿ Stiff Super Grip II Forceps; FORCEPS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Katalyst Surgical, LLC 27ga DEX¿ Stiff Super Grip II Forceps; FORCEPS, OPHTHALMIC Back to Search Results
Model Number DVF4036-27
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/27/2025
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
27ga DEX¿ Stiff Super Grip II Forceps
Common Device Name
FORCEPS, OPHTHALMIC
Manufacturer (Section D)
Katalyst Surgical, LLC
722 goddard avenue
chesterfield MO 63005
Manufacturer (Section G)
Katalyst Surgical, LLC
722 goddard avenue
chesterfield MO 63005
Manufacturer Contact
meryl wilhelm
722 goddard avenue
chesterfield, MO 63005
+1(636)536-5950
MDR Report Key50237258
Report Number3007589150-2026-00013
Device Sequence Number21426113
Product Code HNR
UDI-Device Identifier10840096202608
UDI-Public10840096202608
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDVF4036-27
Device Catalogue NumberDVF4036-27
Device Lot NumberUnknown
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/07/2025
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 03/27/2025
Initial Report FDA Received Date08/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-