| Model Number |
0675 |
| Medical Device Problem Code |
High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this recently implanted right ventricular (rv) lead exhibited high capture thresholds.Technical services (ts) was consulted and determined the measurements rose overnight and due to the patient being dependent, the physician and local field representative opted to reposition the lead.Ts discussed programming the system to the left ventricular only but advised against it as the rv port would remain active and could cause oversensing of signals and reset timing cycles.Ts recommended using cautery mode due to the system being biventricular.No additional adverse patient effects were reported.This rv lead remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.The lead was not returned for analysis; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|