Model Number 693565
Medical Device Problem Codes Over-Sensing (1438); Pacing Problem (1439); Defibrillation/Stimulation Problem (1573); Under-Sensing (1661)
Health Effect - Clinical Code Tachycardia (2095)
Date of Event 04/30/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) may have inappropriately classified ventricular tachycardia (vt) episodes as supra ventricular tachycardia (svt) and delayed therapy due to stability and onset discriminators.It was also reported that there was right ventricular (rv) lead oversensing and undersensing observed which may have resulted in over pacing and delayed episode detection at times.A lead integrity alert (lia) triggered for the rv lead due to high rate non-sustained episodes and elevated sensing integrity counter (sic).The crt-d and rv lead remain in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10:  511212 lead implanted: (b)(6) 2009 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sprint Quattro Secure S MRI SureScan
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50237828
Report Number2649622-2026-23868
Device Sequence Number18706446
Product Code LWS
UDI-Device Identifier00613994446350
UDI-Public(01)00613994446350(17)110107
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date01/07/2011
Device Model Number693565
Device Catalogue Number693565
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2026
Initial Report FDA Received Date08/13/2026
Date Device Manufactured01/15/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
672625 LEAD, DTPA2D1 CRT-D
Outcome Attributed to Adverse Event Life Threatening;
Patient Age77 YR
Patient SexMale
Patient Weight99 LB
-
-