• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Respironics, Inc. Trilogy 100; Ventilator, Continuous, Facility Use

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Respironics, Inc. Trilogy 100; Ventilator, Continuous, Facility Use Back to Search Results
Model Number 1054260
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
A trilogy 100 ventilator, u.S.A.Device was returned to the manufacturer for evaluation with no initial alleged complaint.During the evaluation at the manufacturer's product investigation laboratory (pil), the service technician confirmed that a thermal event occurred but could not determine the cause of the thermal event.Additionally, the device sn/mn label had to be added since the original had the incorrect gtin/revision.The device is being forwarded back to the customer via service.A final report is being filed, if additional information is received that changes the outcome of the investigation and associated coding, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The reported failure of thermal issue is accommodated in the product risk management file as being linked to hazard with description fire, flame, smoke.Complaints for this failure were reviewed to assess for the observed probability levels and compare with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Trilogy 100
Common Device Name
Ventilator, Continuous, Facility Use
Manufacturer (Section D)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key50239311
Report Number2518422-2026-043178
Device Sequence Number10296894
Product Code CBK
UDI-Device Identifier00606959015364
UDI-Public(01)00606959015364
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 08/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1054260
Device Catalogue Number1054260
Device Lot Number150415
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2026
Initial Report FDA Received Date08/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/23/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-