| Model Number |
1054260 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A trilogy 100 ventilator, u.S.A.Device was returned to the manufacturer for evaluation with no initial alleged complaint.During the evaluation at the manufacturer's product investigation laboratory (pil), the service technician confirmed that a thermal event occurred but could not determine the cause of the thermal event.Additionally, the device sn/mn label had to be added since the original had the incorrect gtin/revision.The device is being forwarded back to the customer via service.A final report is being filed, if additional information is received that changes the outcome of the investigation and associated coding, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of thermal issue is accommodated in the product risk management file as being linked to hazard with description fire, flame, smoke.Complaints for this failure were reviewed to assess for the observed probability levels and compare with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
|
| |
|
Search Alerts/Recalls
|
|
|