| Model Number |
VXSK0121 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pericardial Effusion (3271)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during atrial fibrillation procedure a farawave and versacross access solution were selected for use.When trying to perform basket in the ripv, the spline became inverted.It was tried manually fixing the spline, but after going back into the left atrium, the catheter inverted again.Manually fixing the splines but did not work.Ablation was finished and it was noticed a pericardial effusion around 4 hours after the procedure.The catheter had some resistance with the wire when trying to take the catheter out of the body and the splines were inverted at that point.Patient was admitted to the hospital and a pericardiocentesis was performed.Patient has been discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
We are unable to provide the complete unique identifier (udi) at the time of the submission of this report, as it was not yet available.If additional details become available, a supplemental report will be submitted.
|
| |
|