| Model Number |
9-ACP2-010-031 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 31mm amplatzer amulet left atrial appendage occluder was implanted in the patient.The implanted amulet was embolized that evening.A transcatheter retrieval was performed but, unsuccessful so surgery is being considered.The amulet was currently in the left ventricle apex.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 31mm amplatzer amulet left atrial appendage occluder was implanted in the patient.The sizing of the device was determined by feops.During procedure, two different sized devices were attempted and this was the third attempt of the second size.The left atrial pressure was 10mmgh initially so saline was given, later over 12mmhg.The amulet was implanted after satisfying close criteria and tension test was performed.The shape of the device was marshmallow or slightly above, not a roman helmet.The implanted amulet was embolized that evening.A transcatheter retrieval was performed but, unsuccessful so surgery is being considered.The amulet was currently in the left ventricle apex.Later the device got removed and the patient was reported recovering.
|
| |
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of migration or expulsion of device (device embolization) was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on the information received, the cause of the reported device embolization cannot be conclusively determined.There were no complaints associated with any other devices from the lot.The reported that the unexpected medical intervention and surgical intervention results from case-specific circumstances, as the device was snared and explanted the device to address the embolization.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|