It was reported that the staplers keep jamming.A photo and a sample were received for investigation; the account provided a photo of two staplers, and the two opened and unused skin staplers were returned for review and functionally tested by releasing staples into a skin pad.There were no visual defects noted with the component, neither stapler had jammed staples when received, and both staplers punched the staples into the skin pad securely, without jamming; the reported issue was unable to be replicated and the cause code was not confirmed with received sample.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|