An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse replaced integrated electrosurgical unit (iesu) to resolve the issue.The system was verified and ready to use.The generator returned for failure analysis with reported problem and the issue was confirmed and replicated.The reported issue was confirmed via logs, with errors u-02 and m-0 occurring on june 21,2026.Visual inspection found no issues related to the reported event.Fa inspection was able to replicate the reported event; unit was tested on system, error u-02 occurs on startup and remains stuck on splash screen, preventing access to the logs.The probable cause of error was attributed to a faulty electrical component inside the generator.
|