• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic Back to Search Results
Model Number 72404133
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code Urinary Incontinence (4572)
Date of Event 07/16/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this artificial urinary sphincter (aus) experienced recurring urinary incontinence and the device had a size issue.As a result, the device was explanted and replaced.A smaller cuff was implanted.No further patient complications were reported.
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications prior to distribution.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the events of urinary incontinence and inadequate cuff size/loose cuff were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.Model number/catalog number: 72400024, batch/lot number:1100856197, model/catalog description: balloon fgs 61-70 cm, unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127, batch/lot number: 1100801479, model/catalog description: control pump fgs iz, unique identifier (udi) #: (b)(4).Correction to f10 impact codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 Artificial Urinary Sphincter
Common Device Name
Device incontinence mechanical/hydraulic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50243516
Report Number2124215-2026-43549
Device Sequence Number13329040
Product Code EZY
UDI-Device Identifier00878953003115
UDI-Public(01)00878953003115(17)270227(10)1100686118
Combination Product (Y/N)Y
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72404133
Device Catalogue Number72404133
Device Lot Number1100686118
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2026
Supplement Date Received by Manufacturer08/21/2026
Initial Report FDA Received Date08/13/2026
Supplement Report FDA Received Date09/18/2026
Date Device Manufactured02/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age79 YR
Patient SexMale
-
-