It was reported that the ventilator alarmed, the tidal volume had decreased significantly, and the tidal volume fluctuated widely.Gas was observed being exhaled from the patient¿s mouth, inflating the cuff was ineffective, and the cuff pressure was found to be significantly low.External factors such as circuit disconnection, water accumulation, and air trapping were ruled out, and cuff leakage was considered.The ventilator oxygen concentration was adjusted to 100%, secretions and saliva were fully suctioned, the original endotracheal tube was removed and replaced, and the procedure went smoothly.After reconnection to the ventilator, the alarm disappeared, the original ventilator oxygen concentration was restored, and after 5 minutes of observation, oxygen saturation was 100%.
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|