• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Dexcom, Inc. Dexcom G7 15 Day Continuous Glucose Monitoring System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Dexcom, Inc. Dexcom G7 15 Day Continuous Glucose Monitoring System Back to Search Results
Catalog Number SW13355
Medical Device Problem Code Unintended Application Program Shut Down (4032)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a dexcom app crash occurred.Data was evaluated and the allegation was confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Dexcom G7 15 Day Continuous Glucose Monitoring System
Common Device Name
Continuous Glucose Monitor
Manufacturer (Section D)
Dexcom, Inc.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
jaime chavez
6340 sequence dr.
san diego, CA 92121
+1(858)200-0200
MDR Report Key50244594
Report Number3004753838-2026-224607
Device Sequence Number9883120
Product Code QBJ
UDI-Device Identifier00386270005168
UDI-Public(01)00386270005168
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K213919
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 08/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberSW13355
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/29/2026
Initial Report FDA Received Date08/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
Patient Weight195 LB
-
-