Block h6:
device code a0401 is being used to capture the reportable event of suture break.Block h11 device evaluation:
the device was not returned for analysis; therefore, a technical analysis could not be performed.For this reason and due to the lack of evidence is unable to confirm the reported event.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.A similar complaint review was completed.There were no emerging trends identified that warrant further action.Based on all the gathered information, there is not enough evidence to determine whether the reported event was due to the physician's technique during the procedure or was related to a device malfunction.Additionally, for the event additional device required, it happened during the procedure, and the most probable cause of this reported issue cannot be established due to lack of evidence.Without proper evaluation of the device, it remains unknown the most probable causes that contributed to the event.Also, the evidence from the product record review did not identify a potential product quality issue or new patient harm.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of "cause not established".
|