Model number/catalog number: 72400024,
serial number: n/a,
batch/lot number: 1100609858,
model/catalog description: balloon fgs 61-70 cm,
unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127,
serial number: n/a,
batch/lot number: 1100635431,
model/catalog description: control pump fgs iz,
unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.There was no report of a device performance allegation.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the events of atrophy and urinary incontinence were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of adverse event related to patient condition was assigned to this investigation, as the patients condition, disease, or anatomy is demonstrably responsible for the adverse event and use of the device has neither caused nor otherwise influenced the adverse event.
|