| Model Number |
D121 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Syncope/Fainting (4411)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this implantable cardioverter defibrillator (icd) system visited the hospital and reported experiencing a syncopal episode.The device was interrogated and the health care professional (hcp) called technical services (ts) for a consult review of the presenting electrogram (egm).Upon review, the egm showed device recorded a ventricular episode that correlated with the syncope episode.Further review of the ventricular episode showed that patient was in a ventricular tachycardia (vt) with the ventricular rate greater than atrial rate.The device delivered two shocks and successfully converted the rhythm.No adverse patient effects were reported.This icd remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|