• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; Implantable Device

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; Implantable Device Back to Search Results
Model Number U125
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/15/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient developed an infection of the cardiac resynchronization therapy pacemaker (crt-p) system and underwent an explant procedure of all components on (b)(6) 2026.The procedure started with extraction of the lead that was placed for left bundle branch area pacing.Minimal laser was required and the lead was successfully removed.Next, the atrial lead was extracted and during removal a loss of capture occurred on the temporary right ventricular (rv) pacing lead due to the lead catheter being pulled on.The patient was noted to be dependent, with no underlying rhythm.The healthcare provider (hcp) moved to adjust the temporary pacing device and inadvertently caught his foot on the laser equipment and pulled on the atrial lead, dislodging it from the heart.At the same time another hcp was able to increase the output on the temporary device.After the atrial lead was extracted the patient's blood pressure began to drop when capture was lost.Medication was administered and the patient underwent a transesophageal echocardiogram (tee) to examine the heart and no issues were observed.Once the patient's bp was stabilized, and capture with the temporary rv lead was confirmed, the existing right ventricular (rv) lead was extracted.The extraction was successful; however, during removal the laser lead catheter was hooking onto the temporary pacing lead and so it was decided to add another temporary pacing lead for patient safety.After several minutes the device pocket area was closed, and the team began to prepare the patient for a semi-permanent pacemaker and the patient's bp dropped into the 40's.The cardiothoracic surgical team was called stat, and the staff began chest compressions.The patient's chest was opened but ultimately the patient did not survive.All system components were sent to the hospital pathology department per protocol and thus won't be returned.
 
Additional Manufacturer Narrative
Investigation summary/conclusion: with the available information, boston scientific concluded the clinical observation of infection is a known inherent risk of the device and is noted within the product's instructions for use (ifu), as well as the product's risk documentation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE CRT-P
Common Device Name
Implantable Device
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
+1(651)582-6168
MDR Report Key50248282
Report Number2124215-2026-43664
Device Sequence Number15970216
Product Code NKE
UDI-Device Identifier00802526559389
UDI-Public(01)00802526559389(17)250209
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P030005
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/09/2025
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number725077
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2026
Initial Report FDA Received Date08/14/2026
Date Device Manufactured02/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-