H3: product analysis: evaluation determined that distal bearings are detached.The likely cause of failure could not be determined.It was also noted that the burr cannot be locked, the unlock/lock ring does not turn, the unlock/off lock ring is worn, the spindle housing is worn, the drive shafts are pitting, some parts could not be disassembled due to corrosion, the device was heavily polluted internally, the colour ring is faded, and the laser markings are faded.This regulatory report is being submitted due to retrospective review through capa 672570.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|