MEDTRONIC NAVIGATION, INC. PLASMABLADE 4.0; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
|
Back to Search Results |
|
| Model Number |
PS200-040 |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
08/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from a healthcare provider regarding a handpiece.It was reported that during pacemaker replacement surgery the surgical site was re-prepped with an alcohol-based chlorhexidine solution following patient movement and difficulty securing iv access, which required multiple interventions.A significant amount of the antiseptic solution migrated beneath the surgical drape and saturated the underlying draping material.The handpiece was then activated before the alcohol vapors had fully dissipated, resulting in ignition and a fire beneath the drape.The fire was extinguished within approximately one minute; however, the patient sustained second- and third-degree burns.No product defect was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3): no parts have been received by the manufacturer for evaluation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|