Datascope Corp. TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0684-00-0608 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Other
|
|
Event or Problem Description
|
|
It was reported that during preparation for insertion of a transray plus 40cc intra-aortic balloon (iab) catheter, negative pressure was maintained on the balloon using the supplied one-way valve and a 30 ml syringe while a manual flush was performed through the guidewire inner lumen.During the flush, flushing solution was observed leaking into the balloon interior.The event occurred before the iab catheter was fully inserted into the patient.The affected catheter was not used, and iabp therapy was successfully completed using another new transray plus 40cc catheter.The intended insertion site was the right femoral artery, and the aortic condition was reported as normal.The guidewire was manufactured by getinge; the dilator and sheath were manufactured by another company.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated field: h2, g6, h11, b5, b6, a1, d9, h3.Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|