Model Number 0684-00-0576-01U
Medical Device Problem Codes Leak/Splash (1354); Gas/Air Leak (2946); Material Deformation (2976); Moisture or Humidity Problem (2986)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: b4, g3, g6, h2, h3, h6 (medical device problem code, type of investigation, investigation findings, component codes and investigation conclusions), h11 corrected fields: d10, d4 serial (revert this field to blank, as it was inadvertently submitted initially) and d8 (revert this field to blank, as it was inadvertently submitted initially), e2, g2.The iab was returned with the membrane completely unfolded and blood on the exterior of the catheter and between the catheter and the sheath.The sheath was not a maquet product.The extender and pressure tubing were also returned.A photo was provided and reviewed.Three kinks were observed on the catheter tubing approximately 37.1cm, 37.3cm and 57.9cm from iab tip.Additionally visual examination did not identify any condensation inside the returned catheter.An underwater leak test of the balloon, catheter, y fitting, extracorporeal, extender and pressure tubing was performed and no leaks were detected.The iab was placed on the cardiosave pump and the iab fully inflated and deflated.The iab was then pumped for two hours which represents one complete autofill cycle.The iab pumped normally and no alarm sounded from the pump.The evaluation determined kinks in the catheter tubing.It is difficult to determine when or how a kink in the catheter occurs.Although we did not repeat the event in the laboratory setting, a kink in the catheter can cause inflation difficulty or an alarm.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend.Based on the rational provided above, no escalation to the capa process is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sensation Plus 8Fr. 50cc IAB with Accessories, US Only
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50252037
Report Number2248146-2026-0001402
Device Sequence Number10008244
Product Code DSP
UDI-Device Identifier10607567109633
UDI-Public(01)10607567109633(10)3000520954(17)281219(11)251219
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K133074
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0576-01U
Device Catalogue Number0684-00-0576-01U
Device Lot Number3000520954
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/27/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2026
Supplement Date Received by Manufacturer09/22/2026
Initial Report FDA Received Date08/14/2026
Supplement Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Cardiosave
Patient SexUnknown
-
-