| Model Number |
102901 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Codes |
Undesired Nerve Stimulation (1980); Inadequate Pain Relief (2388)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient implanted with an evoke spinal cord stimulation (scs) system reported undesirable increase in stimulation following an abdominal surgery.Electrocautery use was reported during the procedure.During an impedance check, the patient reported unintended occurrence of stimulation.Increased artifacts were observed, and difficulty obtaining ecap was noted during the reprogramming attempt.These observations did not exist during visits prior to the abdominal surgery.A revision procedure was performed and the evoke closed loop stimulator (cls) was replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of this event is in progress.Once the investigation is complete, a supplemental report will be submitted.
|
| |
|