Saluda Medical PTY LTD Evoke Closed Loop Stimulator (CLS) (US); SCS IPG Back to Search Results
Model Number 102901
Medical Device Problem Codes Signal Artifact/Noise (1036); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Codes Undesired Nerve Stimulation (1980); Inadequate Pain Relief (2388)
Date of Event 07/21/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A patient implanted with an evoke spinal cord stimulation (scs) system reported undesirable increase in stimulation following an abdominal surgery.Electrocautery use was reported during the procedure.During an impedance check, the patient reported unintended occurrence of stimulation.Increased artifacts were observed, and difficulty obtaining ecap was noted during the reprogramming attempt.These observations did not exist during visits prior to the abdominal surgery.A revision procedure was performed and the evoke closed loop stimulator (cls) was replaced.
 
Additional Manufacturer Narrative
Investigation of this event is in progress.Once the investigation is complete, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Evoke Closed Loop Stimulator (CLS) (US)
Common Device Name
SCS IPG
Manufacturer (Section D)
Saluda Medical PTY LTD
5 eden park drive
macquarie park, nsw 2113
AS  2113
Manufacturer (Section G)
Saluda Medical Pty Ltd
5 eden park drive
macquarie park, nsw 2113
AS   2113
Manufacturer Contact
sadie mauser
5 eden park drive
macquarie park, nsw 2113
AS   2113
MDR Report Key50252185
Report Number3021836309-2026-00238
Device Sequence Number18687803
Product Code LGW
UDI-Device Identifier09352307001572
UDI-Public09352307001572
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P190002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number102901
Device Catalogue Number3042
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/21/2026
Supplement Date Received by Manufacturer08/25/2026
Initial Report FDA Received Date08/14/2026
Supplement Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-