Datascope Corp. CS100 Intra-Aortic Balloon Pump, Chinese, 220V; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0998-00-3013-45
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Due to characterization limit e1: event site telephone: (b)(6) updated fields: b4, b5, d9, e1(initial reporter, event site telephone), g3, g6, h2, h3, h6(type of investigation, investigation findings, component codes, investigation conclusions), h11.The fault indicates a problem with the front end board, and replacement is needed.No repairs have been performed so far.The fse reported that the hospital is currently going through the necessary procedures to decide whether to proceed with the recommended repairs.If further updates will be provided, this investigation will be reopened and updated with new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 Intra-Aortic Balloon Pump, Chinese, 220V
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
ewelina chorlowska
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key50252576
Report Number2249723-2026-0003853
Device Sequence Number18629764
Product Code DSP
UDI-Device Identifier10607567107790
UDI-Public(01)10607567107790
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K031636
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-3013-45
Device Catalogue Number0998-00-3013-45
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2026
Supplement Date Received by Manufacturer08/28/2026
Initial Report FDA Received Date08/14/2026
Supplement Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1