• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Zimmer Surgical, Inc. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Zimmer Surgical, Inc. ZIMMER AIR DERMATOME Back to Search Results
Catalog Number 00880100100
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Iit was reported that outside of surgery the unit was sent in for preventative maintenance and later discovered to need repair.Investigation results found device had inconsistent speed.There was no patient involvement.Due diligence is complete.
 
Additional Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the device came for pm.Few components were damaged and worn out.Defective motor and control bar were loose.Spring seal was damaged.Control shaft torque was 1.Ball plunger was not in the correct position.Dowel pins were not flush.Adj.Cams were not flush.All worn or damaged components were replaced to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.Able to confirm reported event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Common Device Name
Dermatome
Manufacturer (Section D)
Zimmer Surgical, Inc.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
Zimmer Surgical, Inc.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
+1(574)526-0384
MDR Report Key50252680
Report Number0001526350-2026-00982
Device Sequence Number13291678
Product Code GFD
UDI-Device Identifier00889024375901
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00880100100
Device Lot Number64187488
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/11/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2026
Initial Report FDA Received Date08/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-