| Catalog Number |
00880100100 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Iit was reported that outside of surgery the unit was sent in for preventative maintenance and later discovered to need repair.Investigation results found device had inconsistent speed.There was no patient involvement.Due diligence is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the device came for pm.Few components were damaged and worn out.Defective motor and control bar were loose.Spring seal was damaged.Control shaft torque was 1.Ball plunger was not in the correct position.Dowel pins were not flush.Adj.Cams were not flush.All worn or damaged components were replaced to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.Able to confirm reported event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|