Datascope Corp. CS100 Intra-Aortic Balloon Pump, English, 110V; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0998-00-3013-53 |
| Medical Device Problem Codes |
Device Alarm System (1012); Inability to Auto-Fill (1044); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that, the cs100 intra-aortic balloon pump (iabp) notified alarms but no alarm codes given.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that no alarm codes were given.Fse was not able reproduce the issue.During evaluation, gas loss alarm was also reported.Fse was not able to find any unit issues.Completed repair successfully.Product passed all functional and safety tests per manufacturer specifications.Returned unit to customer for use.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: h2, h11, g6, h6, a1, b5.Corrected fields: g2.Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|