It was reported that during an unknown procedure, it was observed that the device did not respond after firing.A new psee45a was subsequently used, and the procedure was completed successfully.There was no adverse patient consequence nor surgical delay reported.No additional information provided.
|
|
(b)(4).Date sent: 8/14/2026.D4: batch #786c37.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent:
was there any difficulty opening the device jaws?
if yes, what steps were taken to open the jaws?
if the jaws could not be opened, how was the device removed?
was the firing sequence started but not completed?
what was the appearance of the deployed staples (b-formed, malformed, goal post/legs straight)?
were there staples missing from the staple line?
investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one psee45a device was returned with no apparent damage and with no reload present.The device was returned with a non-working firing mechanism.No further functional test could be performed due to the condition of the device.In order to verify the condition of the internal components, the device was disassembled.Upon disassembling, the pcb board was noted to be non-functional.No conclusion could be reached as to what may have caused the pcb board to be damaged.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch
as part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch number, c9d84w, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 786c37, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|