• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Mexico S. De R.L. De CV EVOLUT FX PLUS VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Mexico S. De R.L. De CV EVOLUT FX PLUS VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number EVFXPLUS-26
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Insufficient Device Problem Information (3190); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Bradycardia (1751); Tachycardia (2095)
Date of Event 07/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Additional information was received that the valve was implanted into a patient with severe aortic insufficiency.The first valve dislodged ventricular and was snared up into the ascending aorta.Subsequently a second valve was implanted.
 
Event or Problem Description
It was reported that following the implant of a transcatheter aortic valve, a second valve was implanted valve-in-valve for an unspecified reason.An id card was sent with the serial number of the valve that was initially implanted, and replaced.Subsequently, the serial number was corrected, and a new id card with a data correction letter was sent.Approximately 6 days later, a permanent pacemaker was implanted for brady/tachy syndrome.
 
Additional Manufacturer Narrative
Updated a.4, b.5 and h.6.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT FX PLUS VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
Medtronic Mexico S. De R.L. De CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
Medtronic Mexico S. De R.L. De CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
+353(091)708096
MDR Report Key50253405
Report Number9617601-2026-05135
Device Sequence Number18659680
Product Code NPT
UDI-Device Identifier00763000716134
UDI-Public(01)00763000716134(17)280325(20)01
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberEVFXPLUS-26
Device Catalogue NumberEVFXPLUS-26
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2026
Supplement Date Received by Manufacturer08/14/2026
Initial Report FDA Received Date08/14/2026
Supplement Report FDA Received Date09/11/2026
Date Device Manufactured03/26/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
Patient Weight121 LB
-
-