| Model Number |
A209 |
| Medical Device Problem Codes |
No Audible Alarm (1019); Interrogation Problem (4017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This supplemental is being submitted to correct the tab b.Adverse event/product prob section for the adverse event/product problem.The codes should be product problem and serious injury.The outcomes attrib to adv event code should be hospitalization - initial or prolonged.This supplemental report is being filed to provide additional information that the product has been explanted and returned.There were no additional adverse patient effects.It was reported that this device could not be interrogated.Magnet application over the device did not yield tones.The device has been implanted greater than ten years and the patient has not been compliant with follow-ups.Technical services suspects the battery is dead and discussed some options, such as a life vest.There were no adverse patient effects.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental is being submitted to correct the tab b.Adverse event/product prob section for the adverse event/product problem.The codes should be product problem and serious injury.The outcomes attrib to adv event code should be hospitalization - initial or prolonged.
|
| |
|