Catalog Number SXPP1A400
Medical Device Problem Codes Break (1069); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a laparoscopic surgery on (b)(6) 2026 and a barbed suture was used on the uterine muscle layer.During the procedure, the surgeon inserted the suture into the abdominal cavity through the puncture hole and found that the problem of pulling off suture needle happened and the tip of the needle was hooked, making it unusable.The suture was immediately removed and replaced with a new one.This incident did not cause substantial harm to the patient, but delayed surgical hemostasis, which affected surgical efficiency.There were no adverse patient consequences.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was requested but unavailable: ¿ when was the suture break (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify ¿ were there any patient consequences? ¿ was the needle piece(s) retrieved during the same procedure? ¿ were x-rays taken to locate the needle piece(s)? ¿ what measures were taken to retrieve the broken piece? ¿ was there any additional tissue damage as a result of searching for the needle piece? ¿ if not retrieved, in what tissue structure the broken needle was retained? is there any plan in place to remove the needle piece in the future? h6 investigation findings: c22 - photo review h3 investigation summary: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following observations were made: the corresponding lot number and product code is not visible in the image.An apparent needle detachment is observed.Due to the orientation and positioning of the needle within the photograph, a conclusive assessment of the failure endpoint and the needle attachment hole could not be performed.Consequently, the exact mode and location of failure could not be definitively determined based on the available visual evidence.As the device was not returned, an analysis investigation could not be performed.Based on the photo review, the event reported is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the device was not returned our evaluation is limited.We value the opportunity to fully analyze the device upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
Suture, surgical, absorbable, polydioxanone
Manufacturer (Section D)
Ethicon Inc.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
Ethicon Inc.-Juarez
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
+1(303)552-6892
MDR Report Key50255021
Report Number2210968-2026-08151
Device Sequence Number21374773
Product Code NEW
UDI-Device Identifier10705031219366
UDI-Public(01)10705031219366(17)270731(10)10A36G
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K182873
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Type of Report Initial
Report Date (Section B) 08/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP1A400
Device Lot Number10A36G
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2026
Initial Report FDA Received Date08/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-