| Model Number |
VXSK0122 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cardiac Tamponade (2226); Cardiac Perforation (2513); Pericardial Effusion (3271)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that the patient experienced cardiac tamponade, pericardial effusion, and perforation of the left atrial appendage.The patient died.During a pulse field ablation (pfa) procedure to treat paroxysmal atrial fibrillation a versacross access solution radiofrequency (rf) wire was selected for use.A versacross fixed sheath was used with the rf wire to perform the transseptal puncture (tsp).It was noted that the rf wire went across the septum without delivery of rf energy.The physician spent some time checking intracardiac echocardiography (ice) imaging before dilating the septum with the versacross fixed sheath.It was noted that the rf wire did not appear as it normally did going out the left superior pulmonary vein (lspv) on fluoroscopy, though it did not appear entirely abnormal either.The versacross fixed sheath was exchanged for the faradrive.With the farawave catheter inserted past the sheath handle and into the clear shaft, the sheath was aspirated.Only air was aspirated from the sheath; the air was observed being drawn back from the tip of the farawave catheter.The farawave catheter was removed, but air continued to be aspirated.The sheath was exchanged, but the same aspiration issue was observed with the second faradrive once the dilator was removed.This occurred before advancing into the heart.At this time an effusion was noted on ice.The sheath was pulled back, and transthoracic echocardiography was used to confirm the effusion which was located in the left atrial appendage.The procedure was cancelled and pericardiocentesis was performed, then the patient stabilized enough to be brought to the operating room to attempt to patch the perforation.The patient was brought to the intensive care unit (icu) after the operation, though later the bleeding began again, and the patient subsequently died.It was estimated that the perforation was caused by the rf wire after going transseptal.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced cardiac tamponade, pericardial effusion, and perforation of the left atrial appendage.The patient died.During a pulse field ablation (pfa) procedure to treat paroxysmal atrial fibrillation a versacross access solution radiofrequency (rf) wire was selected for use.A versacross fixed sheath was used with the rf wire to perform the transseptal puncture (tsp).It was noted that the rf wire went across the septum without delivery of rf energy.The physician spent some time checking intracardiac echocardiography (ice) imaging before dilating the septum with the versacross fixed sheath.It was noted that the rf wire did not appear as it normally did going out the left superior pulmonary vein (lspv) on fluoroscopy, though it did not appear entirely abnormal either.The versacross fixed sheath was exchanged for the faradrive.With the farawave catheter inserted past the sheath handle and into the clear shaft, the sheath was aspirated.Only air was aspirated from the sheath; the air was observed being drawn back from the tip of the farawave catheter.The farawave catheter was removed, but air continued to be aspirated.The sheath was exchanged, but the same aspiration issue was observed with the second faradrive once the dilator was removed.This occurred before advancing into the heart.At this time an effusion was noted on ice.The sheath was pulled back, and transthoracic echocardiography was used to confirm the effusion which was located in the left atrial appendage.The procedure was cancelled and pericardiocentesis was performed, then the patient stabilized enough to be brought to the operating room to attempt to patch the perforation.The patient was brought to the intensive care unit (icu) after the operation, though later the bleeding began again, and the patient subsequently died.It was estimated that the perforation was caused by the rf wire after going transseptal.
|
| |
|
Additional Manufacturer Narrative
|
|
It was indicated that the device will not be returned for evaluation.When any further relevant information is obtained, a supplemental medwatch will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary
with all the available information, boston scientific is unable to confirm cause of the reported clinical observation of e0605 cardiac tamponade, e0619 pericardial effusion, and e0604 cardiac perforation.The device has not been received for analysis; therefore, a technical analysis of the complaint device could not be completed.Device history record review
it was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review
based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use (ifu).Risk review
a risk review performed for the versacross access solution device confirmed that the events of e0605 cardiac tamponade, e0619 pericardial effusion, and e0604 cardiac perforation are known events defined in the product risk management documentation.These event types have been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion
boston scientific has assigned an investigation conclusion code of no problem detected, known inherent risk of device and supporting evidence that came to this conclusion.Boston scientifics investigation findings were, there were several factors that may have contributed to the adverse event(s), such as excessive force during advancement of the device, physician/scrub tech technique, tortuous or challenging patient anatomy, or prior patient condition.The device met all requirements prior to being approved for final distribution/sale and there was no evidence to suggest that the instructions for use (ifu) was not followed.Correction to section h6 evaluation method codes (disposed).
|
| |
|