Block h6:
imdrf device code a140101 captures the reportable event of balloon failure to deflate.Block h11: investigation results
the returned cre fixed wire dilatation balloon was analyzed, visual and microscopic inspection found that the balloon was detached with a mechanical cut in the inner wire.Also, the distal part with the tip of the balloon and part of the wire were not returned for analysis.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon failure to deflate was not confirmed.The device was returned with the balloon detached and the inner wire with a partially separated section.It is possible that the balloon could not be pulled out of the scope, and the physician had to cut the catheter.It's important to mention that there is evidence of a mechanical cut on the wire and the balloon.Based on all gathered information and the analysis performed, the most probable cause of this complaint is cause not established.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A similar complaint review was completed.There were no emerging trends identified that warrant further action.
|