• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE Fixed Wire; dilator, esophageal

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE Fixed Wire; dilator, esophageal Back to Search Results
Model Number M00558380
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used during an upper endoscopy procedure performed on (b)(6) 2026.It was reported that after dilation, the balloon would not deflate all way down to remove the balloon from the endoscope.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a140101 captures the reportable event of balloon failure to deflate.Block h11: investigation results the returned cre fixed wire dilatation balloon was analyzed, visual and microscopic inspection found that the balloon was detached with a mechanical cut in the inner wire.Also, the distal part with the tip of the balloon and part of the wire were not returned for analysis.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon failure to deflate was not confirmed.The device was returned with the balloon detached and the inner wire with a partially separated section.It is possible that the balloon could not be pulled out of the scope, and the physician had to cut the catheter.It's important to mention that there is evidence of a mechanical cut on the wire and the balloon.Based on all gathered information and the analysis performed, the most probable cause of this complaint is cause not established.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A similar complaint review was completed.There were no emerging trends identified that warrant further action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE Fixed Wire
Common Device Name
dilator, esophageal
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Cork Limited
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key50255895
Report Number3005099803-2026-03336
Device Sequence Number10125488
Product Code KNQ
UDI-Device Identifier08714729202011
UDI-Public(01)08714729202011(17)280129(10)38472803
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00558380
Device Catalogue NumberM00558380
Device Lot Number0038472803
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/06/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2026
Supplement Date Received by Manufacturer08/17/2026
Initial Report FDA Received Date08/14/2026
Supplement Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/29/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-