| Model Number |
3401 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had an inappropriate shock three days post implant due to oversensing of noise.The patient was working underneath the sink when the shock occurred.The sensing vector was set to the secondary vector.The patient also has a pacemaker implanted.The patient had a previous inappropriate shock due to the oversensing of fast pacing.Technical services discussed options with the caller.The clinic elected not to reprogram the device.The system remains implanted.There were no additional adverse patient effects.
|
| |
|