BOSTON SCIENTIFIC CORPORATION LUX-Dx II+?; Insertable Cardiac Monitor Back to Search Results
Model Number M312
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Fluid Discharge (2686)
Date of Event 07/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this insertable cardiac monitor (icm) noticed that there was white liquid in the incision area like pus.The patient was referred to the clinic.The icm remains in service and no adverse patient effects were reported.The icm was explanted due to infection.It was indicated that the incision area was contaminated with mrsa.No new device was implanted.The original icm will not be returned.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of no known device problem was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to procedure.
 
Event or Problem Description
It was reported that the patient with this insertable cardiac monitor (icm) noticed that there was white liquid in the incision area like pus.The patient was referred to the clinic.The icm remains in service and no adverse patient effects were reported.The icm was explanted due to infection.It was indicated that the incision area was contaminated with mrsa.No new device was implanted.The original icm will not be returned.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUX-Dx II+?
Common Device Name
Insertable Cardiac Monitor
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50255921
Report Number2124215-2026-43725
Device Sequence Number10020692
Product Code MXD
UDI-Device Identifier00802526620706
UDI-Public(01)00802526620706(17)271021
Combination Product (Y/N)N
Initial Reporter StatePR
Initial Reporter CountryUS
PMA/510(K) Number
K231328
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial,Followup
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberM312
Device Catalogue NumberM312
Device Lot Number189561
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2026
Supplement Date Received by Manufacturer09/09/2026
Initial Report FDA Received Date08/14/2026
Supplement Report FDA Received Date09/10/2026
Date Device Manufactured04/29/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age52 YR
Patient SexUnknown
-
-