| Model Number |
M312 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this patient application was not connecting.The patient was using the application.There was no response from the insertable cardiac monitor (icm).The patient application software was uninstalled and reinstalled.The icm was not responding to magnet swipe.The patient indicated that they could feel the icm.The patient was referred to the clinic for a device check.The icm remains in service and no adverse patient effects.Additional information indicated that icm was not responding to magnet swipe.They tried to used a donut magnet but there wasn't a response.The representative indicated that they can feel the icm.It was advised that replacement might be needed, and to send the device back for analysis.The icm remains in service and no adverse patient effects were reported.Additional information indicated that this icm was explanted.The icm has been returned to bsc.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product is being evaluated in our post market quality assurance laboratory.This report will be updated when evaluation is complete.
|
| |
|