Model Number M312
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this patient application was not connecting.The patient was using the application.There was no response from the insertable cardiac monitor (icm).The patient application software was uninstalled and reinstalled.The icm was not responding to magnet swipe.The patient indicated that they could feel the icm.The patient was referred to the clinic for a device check.The icm remains in service and no adverse patient effects.Additional information indicated that icm was not responding to magnet swipe.They tried to used a donut magnet but there wasn't a response.The representative indicated that they can feel the icm.It was advised that replacement might be needed, and to send the device back for analysis.The icm remains in service and no adverse patient effects were reported.Additional information indicated that this icm was explanted.The icm has been returned to bsc.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product is being evaluated in our post market quality assurance laboratory.This report will be updated when evaluation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUX-Dx II+?
Common Device Name
Insertable Cardiac Monitor
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
+1(651)582-6168
MDR Report Key50256085
Report Number2124215-2026-43734
Device Sequence Number10246728
Product Code MXD
UDI-Device Identifier00802526620706
UDI-Public(01)00802526620706(17)260315
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K231328
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 08/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/15/2026
Device Model NumberM312
Device Catalogue NumberM312
Device Lot Number128042
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/04/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2026
Initial Report FDA Received Date08/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age35 YR
Patient SexUnknown
-
-