It was reported that during a robotic thoracic procedure, the surgeon attempted to staple the lung tissue with the robotic stapler and it would not clamp.The surgeon then tried an echelon stapler with a black reload and it partially fired.The surgeon continued to try using the echelon staplers to complete the transaction, and ended up opening a total of 6 staplers and at least 15 reloads, but finally completed the transection.There were several staplers with bent anvils, indicating that the tissue was beyond limits for the stapler.No surgical delay or patient consequences were reported.No additional information provided.
|
|
(b)(4).Date sent: 8/14/2026
d4: batch #unk
d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.H4: the device manufacture date is currently not available.Additional information was requested, and the following was obtained:
the products were all discarded.Lot number information is not available.The surgeon cannot remember the sequence of events accurately.He stated that the thickness of the tissue was far too thick for stapling, but had no other options, so he kept trying, and breaking staplers.He finally debrided a pile of staples, cut through the calcified section, and completed the wedge with the stapler.There is no other information available.No lot or batch number was provided; therefore, a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|