• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Impulse Dynamics USA, Inc. OPTIMIZER Smart; Implantable Pulse Generator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Impulse Dynamics USA, Inc. OPTIMIZER Smart; Implantable Pulse Generator Back to Search Results
Model Number CCM X10
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Twiddlers Syndrome (4563)
Date of Event 01/11/2024
Type of Reportable Event Serious Injury
Event or Problem Description
During a normal follow-up appointment, the lead impedance on a patient's implanted os ipg was found to be >2000 on both leads.X-rays were taken and it was confirmed the patient twiddled the device, the second time this had reportedly occurred according to the physician.The physician opted to have the patient's ipg removed.The explanted ipg was scrapped by the hospital and was therefore unavailable for evaluation.However, a dhr review was conducted and no anomalies were found.It is known that twiddling may cause attached ipg leads to become damaged or dislodged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIMIZER Smart
Common Device Name
Implantable Pulse Generator
Manufacturer (Section D)
Impulse Dynamics USA, Inc.
401 route 73 n
bldg 50, ste 100
marlton NJ 08053
Manufacturer Contact
christian toto
401 route 73 n
bldg 50, ste 100
marlton, NJ 08053
+1(856)434-7921
MDR Report Key50258141
Report Number3012563838-2026-00045
Device Sequence Number10125053
Product Code QFV
UDI-Device Identifier00810003380050
UDI-Public00810003380050
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P180036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date03/03/2024
Device Model NumberCCM X10
Device Catalogue Number10-B411-3-XX
Device Lot NumberH4791
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-