This os ipg was explanted on (b)(6) 2026.Previously, the patient reported that during last charging process, the device's battery indicator on the charger screen always remained at 4 bars.The patient continued to perform weekly charging in the last weeks even though they had no feedback from the charger battery indicator lights.An in-clinic follow-up resulted in this same charger behaviour despite the impulse dynamics field representative's use of either the patient's charger or fce charger.The ipg could not be interrogated using any available programmer; no blinking feedback light on the programmer wand turned on, and only the standby light was on.Only a couple of times, the charger was able to communicate with the device for a few seconds, resulting in an error code 2.It was decided by the patient's physician the ipg would be replaced, which occurred on (b)(6) 2026.The explanted ipg is expected to be received by impulse dynamics for further evaluation.
|