| Model Number |
4457 |
| Medical Device Problem Codes |
Pacing Problem (1439); Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Fatigue (1849); Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
10/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) experienced system shortness of breath (sob) and fatigue.It was noted there was suspicious pacing behavior observed on the electrograms (egms).It was hypothesized that at the time of the device upgrade in 2023, the old right ventricular (rv) lead was placed incorrectly in the right atrial (ra) port of this crt-d, and the old ra lead was mistakenly surgically abandoned.The physician performed a pocket exploration and confirmed the old rv lead was attached to the atrial port, and the ra lead was abandoned.The rv lead was removed from the atrial port and replaced with the ra lead.It was also noted the patient would continue to be monitored and that the patient had an increase in ejection fraction since the crt-d was implanted.This rv lead was surgically abandoned.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|