| Model Number |
72404232-10 |
| Medical Device Problem Codes |
Collapse (1099); Fluid/Blood Leak (1250); Inflation Problem (1310); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with an inflatable penile prosthesis (ipp) stated that when he attempted to use the pump last week, the pump completely collapsed, preventing inflation of the device.The patient reported attempting reset techniques; however, these were unsuccessful.A surgical procedure was performed, during which a small hole in the right cylinder was identified, which caused fluid loss.As a result, the ipp was removed and replaced.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404155
serial number: n/a
batch/lot number: 1000303317
model/catalog description: reservoir 65 ml pc/iz
unique identifier (udi) #: (b)(4)
the device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
|
| |
|