Model Number 72404232-10
Medical Device Problem Codes Collapse (1099); Fluid/Blood Leak (1250); Inflation Problem (1310); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with an inflatable penile prosthesis (ipp) stated that when he attempted to use the pump last week, the pump completely collapsed, preventing inflation of the device.The patient reported attempting reset techniques; however, these were unsuccessful.A surgical procedure was performed, during which a small hole in the right cylinder was identified, which caused fluid loss.As a result, the ipp was removed and replaced.No patient complications were reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72404155 serial number: n/a batch/lot number: 1000303317 model/catalog description: reservoir 65 ml pc/iz unique identifier (udi) #: (b)(4) the device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CX MS Pump
Common Device Name
prosthesis, penis, inflatable
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50260538
Report Number2124215-2026-43784
Device Sequence Number18675795
Product Code JCW
UDI-Device Identifier00878953009780
UDI-Public(01)00878953009780(17)211031(10)1000350668
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/31/2021
Device Model Number72404232-10
Device Catalogue Number72404232-10
Device Lot Number1000350668
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2026
Initial Report FDA Received Date08/14/2026
Date Device Manufactured11/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age37 YR
Patient SexMale
-
-