| Model Number |
4088 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) system exhibited high voltage charges after delivering a shock with low out of range impedance measurements.Technical services (ts) was contacted and noted that there appeared to be a short and system replacement was recommended.Tachy therapy was turned off and the patient was fitted for a life vest.Subsequently, this rv lead was explanted and successfully replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|