• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical Amplatzer Septal Occluder; Transcatheter Septal Occluder

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical Amplatzer Septal Occluder; Transcatheter Septal Occluder Back to Search Results
Model Number 9-ASD-015
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/14/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2024, a 15 mm amplatzer septal occluder was implanted into patient.On (b)(6) 2026, it was discovered that the device was embolized into abdominal aorta which was known through echocardiography.
 
Additional Manufacturer Narrative
An event of a device embolization as reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all defined manufacturing specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.Based on the available information, the cause for the reported device embolization could not be determined.There is no indication of a product quality issue with respect to labeling, design, or manufacturing of the device.
 
Event or Problem Description
It was reported that on (b)(6) 2024, a 15 mm amplatzer septal occluder was implanted in the patient using an 8f amplatzer trevisio intravascular delivery system.Intracardiac echocardiography (ice) was used as the imaging modality during the implant procedure.On (b)(6) 2026, during an echocardiographic examination, the device was identified as having embolized to the abdominal aorta.It was reported that the embolized device did not result in a partial or complete obstruction of blood flow.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Amplatzer Septal Occluder
Common Device Name
Transcatheter Septal Occluder
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
St. Jude Medical #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50261413
Report Number2135147-2026-05619
Device Sequence Number10688114
Product Code MLV
UDI-Device Identifier00811806010120
UDI-Public(01)00811806010120(17)260228(10)7887964
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model Number9-ASD-015
Device Catalogue Number9-ASD-015
Device Lot Number7887964
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2026
Supplement Date Received by Manufacturer09/12/2026
Initial Report FDA Received Date08/14/2026
Supplement Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-